Performance of the SRRC scanning photoelectron microscope

Citation
Ih. Hong et al., Performance of the SRRC scanning photoelectron microscope, NUCL INST A, 467, 2001, pp. 905-908
Citations number
7
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
2
Pages
905 - 908
Database
ISI
SICI code
0168-9002(20010721)467:<905:POTSSP>2.0.ZU;2-B
Abstract
A scanning photoelectron microscope has been constructed at SRRC. This SPEM system consists primarily of a Fresnel zone plate (ZP) with an order-selec tion aperture, a flexure scanning stage, a hemispherical electron analyzer. and sample/ZP insertion system. The flexure stage is used to scan the samp le. A hemispherical analyzer with Omni V lens and a 16-channel multichannel detector (MCD) is used to collect photoelectrons. A set of 16 photoelectro n images at different kinetic energies can be simultaneously acquired in on e single scan. The data acquisition system is designed to collect up to 32 images concurrently. including 16 MCD signals, total electron yield and tra nsmitted photon flux. The design and some initial test results of this SPEM station are presented and discussed. (C) 2001 Elsevier Science B.V. All ri ghts reserved.