Refractive X-ray lens for high pressure experiments at SPring-8

Citation
Y. Ohishi et al., Refractive X-ray lens for high pressure experiments at SPring-8, NUCL INST A, 467, 2001, pp. 962-965
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
2
Pages
962 - 965
Database
ISI
SICI code
0168-9002(20010721)467:<962:RXLFHP>2.0.ZU;2-P
Abstract
A stacked compound refractive X-ray lens was designed to produce an efficie ntly focused (phi < 0.1 mm) beam for high-pressure experiments at SPring-8. High-pressure X-ray diffraction requires an intense. high-energy and monoc hromatic X-ray beam in order to penetrate the absorptive window of a diamon d anvil cell (DAC). Our lens, producing a focal spot of 120 x 275 mum(2) an d a peak gain of 12. is well matched to these requirements. It is composed of many plastic chips made by molding. which allows many identical chips to be made precisely. Other advantages of this lens include high throughput. simple energy tunability and easy installation. (C) 2001 Elsevier Science B .V. All rights reserved.