SRS station 16.3: high-resolution applications

Citation
Bm. Murphy et al., SRS station 16.3: high-resolution applications, NUCL INST A, 467, 2001, pp. 1014-1018
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
2
Pages
1014 - 1018
Database
ISI
SICI code
0168-9002(20010721)467:<1014:SS1HA>2.0.ZU;2-C
Abstract
Station 16.3 is a high-resolution X-ray diffraction beamline at Daresbury L aboratory Synchrotron Radiation Source. The data presented demonstrate the high-resolution available on the station utilising the recently commissione d four-reflection Si 1 1 1 monochromator and three-reflection Si 1 1 1 anal yser. For comparison, a reciprocal space map of the two-bounce Si 1 1 1 mon ochromator and two-bounce analyser is also shown. Operation of the station is illustrated with examples for silicon. and for diamond. Lattice paramete r variations were measured with accuracies in the part per million range an d lattice tilts at the are second level (DuMond. Phys. Rev. 52 (1937) 872). (C) 2001 Elsevier Science B.V. All rights reserved.