A superconducting detector endstation for high-resolution energy-dispersive SR-XRF

Citation
S. Friedrich et al., A superconducting detector endstation for high-resolution energy-dispersive SR-XRF, NUCL INST A, 467, 2001, pp. 1117-1120
Citations number
8
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
467
Year of publication
2001
Part
2
Pages
1117 - 1120
Database
ISI
SICI code
0168-9002(20010721)467:<1117:ASDEFH>2.0.ZU;2-3
Abstract
We have built a two-stage adiabatic demagnetization refrigerator (ADR) to o perate cryogenic high-resolution X-ray detectors in synchrotron-based fluor escence applications. The detector is held at the end of a 40 cm cold finge r that extends into a UHV sample chamber. The ADR attains a base temperatur e below 100 mK with about 20 h hold time below 400 mK, and does not require pumping on the liquid He bath. We will discuss cryostat design and perform ance. (C) 2001 Elsevier Science B.V. All rights reserved.