A four-crystal monochromator beamline has been installed by the Physikalisc
h-Technische Bundesanstalt at a bending magnet of the electron storage ring
BESSY II. The monochromatic radiation with very high spectral purity and h
igh spectral resolution in the photon energy range from 1.75 to 10keV is us
ed to calibrate detectors by comparison to a cryogenic electrical substitut
ion radiometer as primary detector standard with relative uncertainties wel
l below 1%, This is one order of magnitude better than all calibrations of
non-energy-dispersive detectors in this spectral range previously performed
. The beamline is also used for the characterization of optical components.
(C) 2001 Elsevier Science B.V. All rights reserved.