Modification of a commercial electrospray nebulizer for operation in a liquid chromatography/mass spectrometry system at flow rates in the low mu L/min range
A. Carrier et J. Parent, Modification of a commercial electrospray nebulizer for operation in a liquid chromatography/mass spectrometry system at flow rates in the low mu L/min range, RAP C MASS, 15(18), 2001, pp. 1681-1684
A simple and inexpensive approach to convert the electrospray nebulizer of
a commercial liquid chromatography/mass spectrometry (LC/MS) system (HP 110
0) to accommodate lower flow rates has been proposed and evaluated. This mo
dification consists of simply replacing the nebulizer needle by a commercia
lly available stainless steel needle with a smaller internal diameter. Expe
riments were conducted in order to optimize operational parameters. Using t
wo different internal diameter needle sizes, flow rates ranging from 1 to 2
50 muL/min could be accommodated. The modification presented allows an exte
nsion of the range of compatible flow rates without major modification of t
he standard design of the interface. Copyright (C) 2001 John Wiley & Sons,
Ltd.