Yq. Ji et al., Determination of traces of Np-237 in environmental samples by ICP-MS afterseparation using TOA extraction chromatography, FRESEN J AN, 371(1), 2001, pp. 49-53
A simple, rapid, cost-efficient, and robust method for separation of Np-237
with an extraction chromatographic column (TOA: tri-n-octylamine on Teflon
powder) is outlined in detail and further improved for direct ICP-MS analy
sis. The column efficiently retained Np-237 in 2 mol L-1 HNO3 medium and al
l of the Np-237 was easily eluted with 0.02 mol L-1 oxalic acid in 0.16 mol
L-1 HNO3 at 95 degreesC. The separated solutions were free from most matri
x elements and were aspirated into the ICP-MS directly. The decontamination
factor for U-238 is more than 10(4). The instrumental detection limit for
Np-237 was 0.46 pg mL(-1), which corresponds to 1.2 x 10(-5) Bq mL(-1). The
method is more rapid than traditional radiometric techniques. It is also c
onsidered to be more suitable for environmental monitoring than existing me
thods based on TOA.