Temperature quantitation and mapping of frozen tissue

Citation
K. Butts et al., Temperature quantitation and mapping of frozen tissue, J MAGN R I, 13(1), 2001, pp. 99-104
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
JOURNAL OF MAGNETIC RESONANCE IMAGING
ISSN journal
10531807 → ACNP
Volume
13
Issue
1
Year of publication
2001
Pages
99 - 104
Database
ISI
SICI code
1053-1807(200101)13:1<99:TQAMOF>2.0.ZU;2-K
Abstract
A method was developed for quantitating the temperature within frozen tissu e with the magnetic resonance (MR) parameter R2*. The pulse sequence uses h alf-pulse excitation and a short spiral readout to achieve echo times as sh ort as 0.2 msec. Fiber-optic temperature sensors were inserted into bovine liver tissue. The tissue was frozen at one end while being held warm at the other end. Once steady state was reached, the parameter R2* was measured. A linear dependence of R2* on temperature was demonstrated. R2* is Independ ent of freeze number and of the orientation of the temperature gradient wit h respect to the main magnetic field. Feasibility in a canine prostate duri ng cryosurgery is demonstrated. (C) 2001 Wiley-Liss, Inc.