Catheter tracking for MR fluoroscopy: Design of a transmit/receive coil for use with a nasogastric tube

Citation
Dj. Herlihy et al., Catheter tracking for MR fluoroscopy: Design of a transmit/receive coil for use with a nasogastric tube, J MAGN R I, 13(1), 2001, pp. 127-130
Citations number
12
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
JOURNAL OF MAGNETIC RESONANCE IMAGING
ISSN journal
10531807 → ACNP
Volume
13
Issue
1
Year of publication
2001
Pages
127 - 130
Database
ISI
SICI code
1053-1807(200101)13:1<127:CTFMFD>2.0.ZU;2-S
Abstract
A coil tuned to 21.3 MHz was incorporated into a nasogastric tube and used as a marker of tube position during magnetic resonance (MR) imaging in a 0. 5-T scanner. Catheter tracking was investigated with the coil used in both transmit/receive and in receive-only modes. Data acquired from this coil we re overlaid on images obtained using the body coil of the scanner. Visualiz ation of the full length of the catheter with local high signal at the tip was achieved with a temporal resolution of approximately 1 second. (C) 2001 Wiley-Liss, Inc.