Improved spatial harmonic selection for SMASH image reconstructions

Citation
Ca. Mckenzie et al., Improved spatial harmonic selection for SMASH image reconstructions, MAGN RES M, 46(4), 2001, pp. 831-836
Citations number
20
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MAGNETIC RESONANCE IN MEDICINE
ISSN journal
07403194 → ACNP
Volume
46
Issue
4
Year of publication
2001
Pages
831 - 836
Database
ISI
SICI code
0740-3194(200110)46:4<831:ISHSFS>2.0.ZU;2-L
Abstract
The fitting of coil sensitivity functions to spatial harmonics is central t o image reconstructions using the simultaneous acquisition of spatial harmo nics (SMASH) technique. It has previously been shown that the selection of the set of spatial harmonics used in a SMASH reconstruction can have a noti ceable effect on the quality of the reconstructed image. However, a mechani sm for automatic selection of the best set of harmonics in any particular s ituation has not been provided. In this work, a modification to the SMASH r econstruction procedure is introduced that allows the use of a weighted ave rage of all possible harmonics in a reconstruction. The new reconstruction procedure is shown to allow automatic selection of the spatial harmonics an d substantially improve SNR for both phantom and in vivo images. (C) 2001 W iley-Liss, Inc.