Further developments in an ac-based capacitance tomography system

Authors
Citation
Wq. Yang, Further developments in an ac-based capacitance tomography system, REV SCI INS, 72(10), 2001, pp. 3902-3907
Citations number
17
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
10
Year of publication
2001
Pages
3902 - 3907
Database
ISI
SICI code
0034-6748(200110)72:10<3902:FDIAAC>2.0.ZU;2-R
Abstract
An ac-based electrical capacitance tomography (ECT) system has been develop ed at UMIST, with sinusoidal excitation and phase-sensitive demodulation (P SD). In this article the overall system design is briefly described and the n discussed in detail are the key design features, needed to achieve a high signal-to-noise ratio and a high data acquisition rate, including transien t process analysis of the first stage, design of an analog Butterworth low- pass filter and optimization of the data acquisition process. The experimen ts show that the ac-based system has signal variance 19 times better than t he previous design, the charge/discharge ECT system, and that a data acquis ition rate of 140 frames per second is achieved for a 12-electrode system. (C) 2001 American Institute of Physics.