Time response of protection in event of vacuum failure based on Nude ionization gauge controller

Citation
H. Gao et al., Time response of protection in event of vacuum failure based on Nude ionization gauge controller, REV SCI INS, 72(10), 2001, pp. 3979-3982
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
72
Issue
10
Year of publication
2001
Pages
3979 - 3982
Database
ISI
SICI code
0034-6748(200110)72:10<3979:TROPIE>2.0.ZU;2-Q
Abstract
This article describes the design and application of fast-response vacuum p rotection sensor module, based on the Nude ionization gauge and a homemade controller named GH07X. A simulative test indicated that the controller's r esponse time was less than 200 mus when 1 atm air rushed into the vacuum sy stem through a pulsed valve with 0.8 mm orifice nozzle and the emitting cur rent of the Nude gauge was 4 mA. The experiment result showed that the resp onse time mainly depended on the gas density as well as the electron emitti ng current of the Nude gauge filament. Compared with the vacuum protection sensors based on sputter ion pump and cold-cathode gauge, GH07X is faster a nd reliable besides, GH07X can be used as an ultrahigh-vacuum slow valve in terlock controller with response time of 100 ms, which is faster than other gauge controllers. The widely used field-bus interface CAN and common seri al interface RS232/RS485 are embedded in GH07X controller system. (C) 2001 American Institute of Physics.