Standardized method comparison for ACS : 180 plus and Immulite sensitive PSA (sPSA) measurement methods

Citation
Y. Kurtulmus et al., Standardized method comparison for ACS : 180 plus and Immulite sensitive PSA (sPSA) measurement methods, ACCRED Q A, 6(9-10), 2001, pp. 424-426
Citations number
7
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ACCREDITATION AND QUALITY ASSURANCE
ISSN journal
09491775 → ACNP
Volume
6
Issue
9-10
Year of publication
2001
Pages
424 - 426
Database
ISI
SICI code
0949-1775(200109)6:9-10<424:SMCFA:>2.0.ZU;2-E
Abstract
The aim of this study was to compare two immunoassay measurement methods, t he ACS: 180 Plus (Chiron Diagnostics) kit and the Immulite sPSA (DPC) kit. Method comparison analysis was performed according to EP9-A; approved guide line of NCCLS 51. Serum samples having a wide range of total prostate-speci fic antigen (PSA) concentrations were evaluated in split-sample analysis. F -test, t-test analysis and regression statistics were performed. In Deming regression analysis the coefficients were as follows; the slope=0.967; y-in tercept=-0.148, r=0.989. An acceptable bias was seen since the systematical error was calculated to have a value less than the total allowable error c alculated from biological variations. Nonparametric evaluation of the area under ROC curves for ACS: 180 Plus and Immulite sPSA were 0.997 and 0.987, respectively. Diagnostic accuracy was at the level of p=0.000 and no statis tical difference was found between the two assay methods.