Y. Kurtulmus et al., Standardized method comparison for ACS : 180 plus and Immulite sensitive PSA (sPSA) measurement methods, ACCRED Q A, 6(9-10), 2001, pp. 424-426
The aim of this study was to compare two immunoassay measurement methods, t
he ACS: 180 Plus (Chiron Diagnostics) kit and the Immulite sPSA (DPC) kit.
Method comparison analysis was performed according to EP9-A; approved guide
line of NCCLS 51. Serum samples having a wide range of total prostate-speci
fic antigen (PSA) concentrations were evaluated in split-sample analysis. F
-test, t-test analysis and regression statistics were performed. In Deming
regression analysis the coefficients were as follows; the slope=0.967; y-in
tercept=-0.148, r=0.989. An acceptable bias was seen since the systematical
error was calculated to have a value less than the total allowable error c
alculated from biological variations. Nonparametric evaluation of the area
under ROC curves for ACS: 180 Plus and Immulite sPSA were 0.997 and 0.987,
respectively. Diagnostic accuracy was at the level of p=0.000 and no statis
tical difference was found between the two assay methods.