A silica-based chiral monolithic column prepared by sol-gel process and che
mical modification of chiral selector was used for enantioseparation of dan
syl amino acids and hydroxy acids by capillary electrochromatography (CEC)
and p-high-performance liquid chromatography (V-HPLC). L-Prolinamide was mo
dified as a chiral selector. The chiral stationary phase (CSP), the chiral
complex of Cu(II) with L-prolinamide, provides an anodic electroosmotic flo
w (EOF) in CEC. The EOF was found to be dependent on applied electric field
strength, the pH, and the composition of mobile phases. Scanning electron
micrograph showed that monolithic columns have the morphology of continuous
skeleton and large through-pore, D-Enantiomers migrated before L-enantiome
rs except for dansyl-(Dns)-DL-Ser. The separation efficiencies of up to 17
600 (D) and 13 200 plates m(-1) (L) were achieved for the separation Of DL-
indole-3-lactic acid.