The thermoluminescent (TL) response, supralinearity function, and relative
efficiency with respect to Co-60 gamma -rays, of TLD-100 exposed to 0.7. 1.
5 and 3 MeV protons have been measured. The TL response of the glow curve p
eaks using two different deconvolutions, separating in either 7 (3-9) or 9
peaks (3-10 and peaks 6a and 6b). was investigated. Results for the suprali
nearity function of peaks 5 and 7-9 and the relative response of peaks 5 an
d 7 are insensitive to the chosen deconvolution scheme. At all energies we
observe that peak 5 is slightly supralinear and that supralinearity increas
es as a function of the peak temperature. The relative efficiencies, measur
ed for the total TL signal (area under the glow curve) and peaks 5 and 7, d
ecrease with the proton energy having values of 0.33, 0.46 and 0.70 (total
TL signal), 0.22, 0.32 and 0.53 (peak 5) and 1.4, 2.6 and 3.6 (peak 7) at 0
.7. 1.5 and 3 MeV, respectively. These data agree with the systematics repo
rted in the literature. The measurements of the supralinearity function are
compared with Monte Carlo track interaction model calculations. The analys
is suggests that the model ion track effective radius and the saturation ra
dius are independent of the incident ion energy. (C) 2001 Elsevier Science
B.V. All rights reserved.