Application of primary plasma standardization to Nd-YAG laser-induced shock wave plasma spectrometry for quantitative analysis of high concentration Au-Ag-Cu alloy

Citation
H. Kurniawan et al., Application of primary plasma standardization to Nd-YAG laser-induced shock wave plasma spectrometry for quantitative analysis of high concentration Au-Ag-Cu alloy, SPECT ACT B, 56(8), 2001, pp. 1407-1417
Citations number
27
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
ISSN journal
05848547 → ACNP
Volume
56
Issue
8
Year of publication
2001
Pages
1407 - 1417
Database
ISI
SICI code
0584-8547(20010831)56:8<1407:AOPPST>2.0.ZU;2-6
Abstract
A study was performed on a laser-induced shock wave plasma generated on hig h concentration Au-Ag-Cu alloys by a Q-switched Nd-YAG laser of 4.8 mJ unde r reduced air pressure of 2 torr. It was found that the total emission inte nsity of the secondary plasma is proportional to the intensity of the prima ry plasma. Assuming linear proportionality between the intensity of the pri mary plasma and the number of atoms vaporized from the target, it is propos ed that quantitative analysis can be applied to the intensities of the anal ytical emission lines normalized by the total intensity of the primary plas ma. This experimental result demonstrated for each metal element shows an e xcellent linear relationship between the normalized emission line intensity and the content of the corresponding element. (C) 2001 Elsevier Science B. V. All rights reserved.