Jp. Holmes et al., Detection of simulated internal tooth resorption using conventional radiography and subtraction imaging, DENTOMAX R, 30(5), 2001, pp. 249-254
Objectives: To compare the diagnostic accuracy of conventional radiography
with digital subtraction imaging in the detection of simulated internal res
orption.
Methods: Simulated resorption cavities were created by the sequential use o
f progressively larger round burs (ISO size range 006-016) in the labial wa
ll of the coronal or apical thirds of the pulp chamber of the maxillary inc
isors of two cadavers. Five viewers examined 120 pairs of radiographs, one
of each pair being baseline (no lesion) and the other with or without a les
ion. The process was then repeated with subtraction images of the pairs of
radiographs. Receiver Operating Characteristic (ROC) analysis was used to c
ompare the diagnostic accuracy of the two imaging modalities.
Results: Using conventional radiography, the minimum lesion size detectable
coronally was in the 'medium' range (ISO size 010 and 012 round burs), but
only 'large' lesions (ISO size 014 and 016) could be detected in the apica
l region. The overall diagnostic accuracy of subtraction imaging was superi
or to conventional radiography but only significant (P <0.05) for coronal l
esions.
Conclusion: Subtraction radiography may be a useful tool for detecting and
monitoring the progress of internal root resorption.