Application of sweeping to micellar electrokinetic chromatography-atmospheric pressure chemical ionization-mass spectrometric analysis of environmental pollutants
K. Isoo et al., Application of sweeping to micellar electrokinetic chromatography-atmospheric pressure chemical ionization-mass spectrometric analysis of environmental pollutants, ELECTROPHOR, 22(16), 2001, pp. 3426-3432
An application study of sweeping, an on-line sample concentration technique
, to micellar electrokinetic chromatography (MEKC) directly combined with m
ass spectrometry (MS) using an atmospheric pressure chemical ionization (AP
Cl) interface, namely MEKC-APCl-MS, was investigated to enhance the concent
ration sensitivity for the analysis of environmental pollutants. Under a ne
utral condition, around 100-fold increase in the concentration sensitivity
was achieved for several aromatic amines and alkyl phthalates as test sampl
es by sweeping-MEKC-APCl-MS compared to conventional MEKC-APCl-MS, whereas
under an acidic condition, 100 to 600-fold sensitivity enhancement was gain
ed for similar solutes. Linearity of the corrected peak area obtained in th
e mass chromatogram against the sample concentration was examined for 3,4-d
ichloroaniline and diisopropyl phthalate (DIPP). The estimated limits of de
tection for 3,4-dichloroaniline and DIPP were 0.6 and 0.4 ppm, respectively
, in terms of the injected sample concentration.