SUCCESSFUL MANAGEMENT OF PALMAZ-SCHATZ STENTS DEFORMED BY INADVERTENTANGIOPLASTY OUTSIDE THE STENTS

Authors
Citation
Tnt. Chu et Fs. Ling, SUCCESSFUL MANAGEMENT OF PALMAZ-SCHATZ STENTS DEFORMED BY INADVERTENTANGIOPLASTY OUTSIDE THE STENTS, Catheterization and cardiovascular diagnosis, 41(4), 1997, pp. 435-439
Citations number
16
Categorie Soggetti
Cardiac & Cardiovascular System
ISSN journal
00986569
Volume
41
Issue
4
Year of publication
1997
Pages
435 - 439
Database
ISI
SICI code
0098-6569(1997)41:4<435:SMOPSD>2.0.ZU;2-#
Abstract
We describe a case in which balloon angioplasty of thrombosed saphenou s vein graft stents resulted in compression and deformation of the ste nts because of inadvertent passage of the guidewire outside the stents , This condition was diagnosed by intravascular ultrasonography, which was then ultimately utilized to direct and optimize placement of addi tional stents to exclude the deformed stents from the vessel lumen. (C ) 1997 Wiley-Liss, Inc.