Analysis of enantiomers giving partially overlapped peaks by using different treatments of the chromatographic ultraviolet signals: quantification ofpseudoephedrine enantiomers
J. Verdu-andres et al., Analysis of enantiomers giving partially overlapped peaks by using different treatments of the chromatographic ultraviolet signals: quantification ofpseudoephedrine enantiomers, J CHROMAT A, 930(1-2), 2001, pp. 95-107
Different strategies for the quantification of partially coeluting optical
isomers have been investigated. The methods tested are based on the use of
different features as the analytical UV signals: peak heights, perpendicula
r drop areas, first and second derivatives of the chromatograms, peak areas
obtained by deconvolution of the overlapped peaks with data fitting optimi
zation, and a multivariate model (principal component regression, PCR). The
amphetamine-derivative drug pseudoephedrine was selected as a model compou
nd. For chromatography, LiChrospher 100 RP18 and a mobile-phase consisting
of methanol and a solution of carboxymethyl-beta -cyclodextrin (the chiral
selector) were used. The UV detector was set at 215 mn. The accuracy obtain
ed with the tested methods at different degrees of overlapping and at diffe
rent concentration ratios between enantiomers was evaluated. The results of
this study demonstrated that the best option for quantification of partial
ly overlapped UV peaks of enantiomers and to obtain the enatiomeric excess
is the use of a PCR model using peak heights, perpendicular drop peak areas
and deconvoluted peak areas as the original variables. The predictive abil
ity of the proposed calibration model is of about 2-8 times better (dependi
ng on the overlapping degree) than that achieved with the other models test
ed. (C) 2001 Elsevier Science B.V. All rights reserved.