STANDARDLESS ANALYSIS IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROSCOPY WITH PD AS MATRIX MODIFIER - COMPARISON OF A NEW ATOMIZATION SYSTEM WITH A COMMERCIAL SYSTEM IN THE DETERMINATION OF PB AND CD IN FRUIT JUICE

Citation
C. Locatelli et al., STANDARDLESS ANALYSIS IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROSCOPY WITH PD AS MATRIX MODIFIER - COMPARISON OF A NEW ATOMIZATION SYSTEM WITH A COMMERCIAL SYSTEM IN THE DETERMINATION OF PB AND CD IN FRUIT JUICE, Annali di chimica, 87(5-6), 1997, pp. 397-404
Citations number
12
Categorie Soggetti
Chemistry Analytical","Environmental Sciences
Journal title
ISSN journal
00034592
Volume
87
Issue
5-6
Year of publication
1997
Pages
397 - 404
Database
ISI
SICI code
0003-4592(1997)87:5-6<397:SAIEAS>2.0.ZU;2-Y
Abstract
A new atomization system, previously proposed for standardless analysi s, is compared with a commercial system in the determination of Pb and Cd in fruit juice with Pd as matrix modifier. It is concluded that, i n addition to an increase in sensitivity, the new atomization system h as a response independent of the matrix. Therefore, no calibration is necessary except for a knowledge of the spectroscopic constant of the line used. Naturally the atomization must be such that all the atoms i njected are simultaneously present in the atomizer. In the commercial system on the other hand, the calibration curve is dependent on the ty pe and concentration of the matrix modifier.