Torsional deformation of an endoscope probe

Citation
Fv. Katopodes et al., Torsional deformation of an endoscope probe, P ROY SOC A, 457(2014), 2001, pp. 2491-2506
Citations number
10
Categorie Soggetti
Multidisciplinary
Journal title
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES
ISSN journal
13645021 → ACNP
Volume
457
Issue
2014
Year of publication
2001
Pages
2491 - 2506
Database
ISI
SICI code
1364-5021(20011008)457:2014<2491:TDOAEP>2.0.ZU;2-F
Abstract
A recent US patent due to Shan describes a method for identifying the locat ion of the colonoscope in the body during endoscopy. The method depends upo n the integration of measurements recorded by a flexible probe inserted in the biopsy channel at the end of a flexible rod. In this paper, classical results for the elastic rod are used to determine the equations required for this integration process. Particular attention i s paid to the effect of initial curvature or non-axisymmetry of the rod. It is shown that these effects cause predictive errors that increase quadrati cally with the angle subtended by the space curve of the channel, and quite stringent limits must be placed on straightness and symmetry if these erro rs are not to be significant.