A solution-based inductively coupled plasma optical emission spectrometric
(ICP-OES) method is described for elemental analysis,vith relative expanded
uncertainties on the order of 0.1% relative. The single-element determinat
ions of 64 different elements are presented, with aggregate performance res
ults for the method and parameters for the determination of each element. T
he performance observed is superior to that previously reported for ICP-OES
, resulting from a suite of technical strategies that exploit the strengths
of contemporary spectrometers, address measurement and sample handling noi
se sources, and permit rugged operation with small uncertainty. Taken toget
her, these strategies constitute high-performance ICP-OES.