Severe wear from retrieved alumina-on-alumina coupled implant: a case report

Citation
S. Affatato et al., Severe wear from retrieved alumina-on-alumina coupled implant: a case report, INT J ARTIF, 24(9), 2001, pp. 655-662
Citations number
30
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology
Journal title
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS
ISSN journal
03913988 → ACNP
Volume
24
Issue
9
Year of publication
2001
Pages
655 - 662
Database
ISI
SICI code
0391-3988(200109)24:9<655:SWFRAC>2.0.ZU;2-L
Abstract
This study details the in vivo wear behavior of an alumina acetabular cup a nd a femoral head on a retrieved non-cemented hip prosthesis. A commercial alumina ceramic-on-ceramic prosthesis was retrieved from a pat ient previously treated for bilateral hip arthrosis in "coxa profunda". Mas sive wear was found on the retrieved alumina ceramic head and acetabular cu p. The total measured penetration depth was 1.9 mm while the total calculat ed weight loss for the acetabular cup was 6.06 g. The study underlines the head-cup instability caused by cup loosening as ma jor cause of severe ceramic wear.