Jr. Gibson et S. Taylor, Numerical investigation of the effect of electrode size on the behaviour of quadrupole mass filters, RAP C MASS, 15(20), 2001, pp. 1960-1964
The behaviour of a quadrupole mass spectrometer (QMS) for variation in rod
size has been investigated using a numerical approach reported previously.
Behaviour is found to vary significantly with the change in rod size and th
e optimum rod size suggested is closer to that used in some commercial QMS
instruments than to the value from analytical approaches. The results allow
an upper limit to manufacturing tolerances to be inferred. The present wor
k also confirms that the size of the QMS housing has only a very small effe
ct on QMS behaviour. Copyright (C) 2001 John Wiley & Sons, Ltd.