Numerical investigation of the effect of electrode size on the behaviour of quadrupole mass filters

Citation
Jr. Gibson et S. Taylor, Numerical investigation of the effect of electrode size on the behaviour of quadrupole mass filters, RAP C MASS, 15(20), 2001, pp. 1960-1964
Citations number
11
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
ISSN journal
09514198 → ACNP
Volume
15
Issue
20
Year of publication
2001
Pages
1960 - 1964
Database
ISI
SICI code
0951-4198(2001)15:20<1960:NIOTEO>2.0.ZU;2-V
Abstract
The behaviour of a quadrupole mass spectrometer (QMS) for variation in rod size has been investigated using a numerical approach reported previously. Behaviour is found to vary significantly with the change in rod size and th e optimum rod size suggested is closer to that used in some commercial QMS instruments than to the value from analytical approaches. The results allow an upper limit to manufacturing tolerances to be inferred. The present wor k also confirms that the size of the QMS housing has only a very small effe ct on QMS behaviour. Copyright (C) 2001 John Wiley & Sons, Ltd.