J. Medved et al., Evaluation of sample pre-treatment procedures for the determination of Cr,Ni and V in biological matrices by ETAAS, TURK J CHEM, 25(3), 2001, pp. 323-331
Electrothermal atomic absorption spectrometry (ETAAS) was applied for the d
etermination of chromium, nickel and vanadium in biological materials (eart
hworms, feather, lucerne). Three different decomposition procedures of biol
ogical matrices are presented. The utilization of NH4H2PO4, Mg(NO3)(2) and
Pd(NO3)(2+) Mg(NO3)(2) as chemical modifiers are investigated and compared.
Optimal temperature conditions for the electrothermal determination of Cr,
Ni and V are selected. Precision of the determination expressed as relativ
e standard deviations for Cr, Ni and V varied from 6 to 22%. The detection
limits are for Cr 8 mug kg(-1), Ni 12 mug kg(-1), and V 36 mug kg(-1). The
accuracy of the measured data were verified by determination of the studied
elements in certified reference materials. The attained analytical results
were in good agreement to each other by all applied methods, as well as to
the certified values. Consequently, the method was applied to biological m
atrices: earthworms, feathers and lucerne which were sampled at the refiner
y immission site.