Determination of zinc in pharmaceutical products by use of a sequential injection analysis system

Citation
Jkf. Van Staden et M. Tsanwani, Determination of zinc in pharmaceutical products by use of a sequential injection analysis system, FRESEN J AN, 371(3), 2001, pp. 376-379
Citations number
23
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
09370633 → ACNP
Volume
371
Issue
3
Year of publication
2001
Pages
376 - 379
Database
ISI
SICI code
0937-0633(200110)371:3<376:DOZIPP>2.0.ZU;2-3
Abstract
A sequential injection analysis (SIA) system is proposed for the determinat ion of zinc in pharmaceutical samples. The method is based on the spectroph otometric detection of zinc using xylenol orange as a colour reagent at 568 nm. The system can be used to monitor zinc at a frequency of 30 samples h( -1) with an average recovery of +/- 98.5% and a relative standard deviation of less than 1%. A linear relationship between peak height and zinc concen tration is obtained between 10 and 60 mg L-1. The results obtained are in g ood agreement with those obtained by use of a standard method.