Aiming for robust high-resolution free respiration coronary artery imaging,
localized tracking volume selective 3D-EPI at spatial resolution 1.25 x 1.
25 x 3 mm and temporal resolution 65 ms was evaluated in 14 normal subjects
. Subject-specific motion tracking factors were measured between diaphragm
navigator and coronary artery in S-1 and A-P directions. Imaging was repeat
ed with and without tracking, accepting 128 cardiac cycles over a 20-mm ran
ge from free breathing. Reference images with minimal respiratory motion (5
-mm rang use LED-guided multiple breathholds. Depending on accurate mo tion
measurement, coronary arteries were imaged with tracking, in a 20-mm range
of free breathing, with increased scan efficiency but without significant
loss of image quality compared to multiple breathhold imaging. (C) 2001 Wil
ey-Liss, Inc.