A method for MR angiography using an RF labeling technique is suggested. Th
e method utilizes a slice-selective spin-lock pulse sequence for tagging th
e spins of inflowing blood. The pulse sequence begins with a spatially sele
ctive 90 degrees (x) RF pulse, followed by a nonselective composite locking
pulse of 135 degrees (y) - n[360 degrees (y)]-135 degrees (y) and by a 90
degrees (-x) pulse. A spoiler gradient is then applied. A rapid imaging sta
ge, which yields a Tp-weighted signal from the tagged spins, completes the
sequence. Untagged spins are thoroughly dephased and consequently suppresse
d in the image. Thus, contrast is obtained without an injection of a contra
st material or image subtraction. Furthermore, the flow of the tagged bolus
can be visualized. The sequence was implemented on phantoms and on human v
olunteers using a 1.5T scanner. The results indicate the feasibility of the
suggested sequence. (C) 2001 Wiley-Liss, Inc.