MR angiography using spin-lock flow tagging

Citation
H. Azhari et al., MR angiography using spin-lock flow tagging, MAGN RES M, 46(5), 2001, pp. 1041-1044
Citations number
16
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MAGNETIC RESONANCE IN MEDICINE
ISSN journal
07403194 → ACNP
Volume
46
Issue
5
Year of publication
2001
Pages
1041 - 1044
Database
ISI
SICI code
0740-3194(200111)46:5<1041:MAUSFT>2.0.ZU;2-I
Abstract
A method for MR angiography using an RF labeling technique is suggested. Th e method utilizes a slice-selective spin-lock pulse sequence for tagging th e spins of inflowing blood. The pulse sequence begins with a spatially sele ctive 90 degrees (x) RF pulse, followed by a nonselective composite locking pulse of 135 degrees (y) - n[360 degrees (y)]-135 degrees (y) and by a 90 degrees (-x) pulse. A spoiler gradient is then applied. A rapid imaging sta ge, which yields a Tp-weighted signal from the tagged spins, completes the sequence. Untagged spins are thoroughly dephased and consequently suppresse d in the image. Thus, contrast is obtained without an injection of a contra st material or image subtraction. Furthermore, the flow of the tagged bolus can be visualized. The sequence was implemented on phantoms and on human v olunteers using a 1.5T scanner. The results indicate the feasibility of the suggested sequence. (C) 2001 Wiley-Liss, Inc.