M. Gaspari et al., Ion trap mass spectrometry as detector for capillary electrochromatographyof peptides: Possibilities and limitations, J MICROCOL, 13(6), 2001, pp. 243-249
Capillary electrochromatography has been coupled to an ion trap mass spectr
ometer via a sheath flow electrospray interface. Peptide analysis was perfo
rmed on 25 cm columns packed with C-18-bonded silica particles, and very fa
st separations were obtained. Full mass spectrometry and tandem mass spectr
ometry (MS/MS) detection were evaluated in terms of sensitivity and scannin
g speed. The separation of seven selected peptides and their detection by M
S/MS in selected reaction monitoring mode was demonstrated. The performance
of this sheath flow interface also was compared with the previously report
ed sheathless interface. Despite the loss in sensitivity (20-40-fold), the
sheath flow interface was shown to be superior in terms of ruggedness, and
allowed the use of higher electric fields to achieve faster analysis times.
(C) 2001 John Wiley & Sons, Inc.