Ion trap mass spectrometry as detector for capillary electrochromatographyof peptides: Possibilities and limitations

Citation
M. Gaspari et al., Ion trap mass spectrometry as detector for capillary electrochromatographyof peptides: Possibilities and limitations, J MICROCOL, 13(6), 2001, pp. 243-249
Citations number
30
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF MICROCOLUMN SEPARATIONS
ISSN journal
10407685 → ACNP
Volume
13
Issue
6
Year of publication
2001
Pages
243 - 249
Database
ISI
SICI code
1040-7685(2001)13:6<243:ITMSAD>2.0.ZU;2-6
Abstract
Capillary electrochromatography has been coupled to an ion trap mass spectr ometer via a sheath flow electrospray interface. Peptide analysis was perfo rmed on 25 cm columns packed with C-18-bonded silica particles, and very fa st separations were obtained. Full mass spectrometry and tandem mass spectr ometry (MS/MS) detection were evaluated in terms of sensitivity and scannin g speed. The separation of seven selected peptides and their detection by M S/MS in selected reaction monitoring mode was demonstrated. The performance of this sheath flow interface also was compared with the previously report ed sheathless interface. Despite the loss in sensitivity (20-40-fold), the sheath flow interface was shown to be superior in terms of ruggedness, and allowed the use of higher electric fields to achieve faster analysis times. (C) 2001 John Wiley & Sons, Inc.