Verification and utilization of sterilizing equipment by dentists in San Luis Potosi, Mexico

Citation
N. Patino-marin et al., Verification and utilization of sterilizing equipment by dentists in San Luis Potosi, Mexico, SALUD PUB M, 43(5), 2001, pp. 455-458
Citations number
8
Categorie Soggetti
Public Health & Health Care Science
Journal title
SALUD PUBLICA DE MEXICO
ISSN journal
00363634 → ACNP
Volume
43
Issue
5
Year of publication
2001
Pages
455 - 458
Database
ISI
SICI code
0036-3634(200109/10)43:5<455:VAUOSE>2.0.ZU;2-J
Abstract
Objective. To assess the utilization of sterilizing equipment used by denti sts, and verification of sterilization using biological indicators. Materia l and methods. A cross-sectional study was conducted in 1999-2000, among 13 0 (65%) dentists having sterilizing equipment, at Facultad de Estomatologia , Universidad Autonoma de San Luis Potosi and Colegio Dental Potosino. Biol ogical indicators for sterilization containing Bacillus subtilis and Bacill us stearothermophilus were used. Results. Thirty autoclaves and 100 dry-hea t sterilizers were evaluated: 23 (17.7%) of them showed bacterial growth.Tw enty-one (16.1%) dentists already were using biological indicators to verif y their sterilizing equipment. Both sterilization methods were found to all ow bacterial growth with similar frequencies (p=>0.66). Conclusions. Few de ntists verify the quality of sterilization process through biological indic ators; bacterial growth and failure of sterilization were evidenced. The En glish version of this paper is available at: http://www.insp.mx/salud/index .html.