L. Modis et al., Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness, CORNEA, 20(7), 2001, pp. 711-714
Purpose. To determine the central corneal thickness values in healthy eyes
with the recently developed Orbscan scanning-slit system, contact and nonco
ntact specular microscopic pachymetry and compare the results to convention
al ultrasonic pachymetry. Methods. In the following sequence, Orbscan, Topc
on SP-2000P noncontact specular microscope, AL-1000 ultrasound, and Tomey c
ontact specular microscope were used to record thickness values. Thirty-fou
r healthy right corneas of 34 healthy subjects were investigated. Results.
Orbscan pachymetry correlated significantly with ultrasound (r = 0.64, p <
0.001), contact (r = 0.45, p < 0.001), and noncontact specular microscopy (
r = 0.72, p < 0.001). Likewise, the Topcon SP-2000P noncontact specular mic
roscopy pachymetry disclosed similar statistical results compared with ultr
asound (r = 0.88, p < 0.001), and contact specular microscopy pachymetry (r
= 0.76, p < 0.001). The mean central corneal thickness results were signif
icantly higher (p less than or equal to 0.01) than ultrasonic values (580 /- 43 mum) using the contact specular microscope (640 +/- 43 mum) or Orbsca
n system (602 +/- 59 mum) but were significantly, lower (p < 0.001) using t
he noncontact specular microscope (547 +/- 49 mum). Conclusions. The result
s indicate that the devices tested cannot be simply used interchangeably. F
or long-term patient follow-up, one specific instrument is recommended. Rec
ently developed pachymetry machines are especially helpful when additional
corneal data such as thickness profile, elevation maps, anterior chamber de
pth, and endothelial morphology are required.