An atmospheric pressure ion lens to improve electrospray ionization at lowsolution flow-rates

Citation
Bb. Schneider et al., An atmospheric pressure ion lens to improve electrospray ionization at lowsolution flow-rates, RAP C MASS, 15(22), 2001, pp. 2168-2175
Citations number
20
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
ISSN journal
09514198 → ACNP
Volume
15
Issue
22
Year of publication
2001
Pages
2168 - 2175
Database
ISI
SICI code
0951-4198(2001)15:22<2168:AAPILT>2.0.ZU;2-W
Abstract
The performance of a reduced flow-rate electrospray ion source is substanti ally improved by the addition of an atmospheric pressure ion lens near the tip of the tapered sprayer. The magnitude and stability of ion signals, det ected within a mass spectrometer, are significantly improved. In addition, the ion lens decreased the dependence of the ion signals on the sprayer pos ition. These benefits are believed to be due to an improvement in the shape of the equipotential lines near the tip of the electrospray ion source. Co pyright (C) 2001 John Wiley & Sons, Ltd.