Background-Defibrillation thresholds (DFT) with standard implantable cardio
verter-defibrillator leads in the right ventricle (RV) may be determined by
weak shock field intensity in the myocardium of the left ventricle (LV). A
dding a shocking electrode in a coronary vein on the middle of the LV free
wall, thereby establishing biventricular defibrillation, substantially redu
ced defibrillation requirements in animals. We investigated the feasibility
of this approach in 24 patients receiving an implantable cardioverter-defi
brillator using a prototype over-the-wire temporary LV defibrillation lead.
Methods and Results-The LV lead was inserted through the coronary sinus, us
ing a guide catheter and guidewire, into a posterior or lateral coronary ve
in whose location was determined by retrograde venography. Paired DFT testi
ng compared a standard system (RV to superior vena cava plus can emulator [
SVC+Can], 60% tilt biphasic shock) to a system including the LV lead. The b
iventricular system was tested with a dual-shock waveform (20% tilt monopha
sic shock from LV --> SVC+Can, then 60% tilt biphasic shock from RV --> SVC
+Can). Twenty patients completed DFT testing. Venography and LV lead insert
ion time was 46 +/- 40 minutes. The biventricular system reduced mean DFT b
y 45% (8.9 +/-1.1 J versus 4.9 +/-0.5 J, P <0.001). Twelve patients (60%) h
ad a standard system DFT :S J, and the biventricular system gave a lower DF
T in all patients. There were no adverse events related to the use of the L
V lead, which was removed after testing.
Conclusions-Internal defibrillation using a transvenously inserted LV lead
is feasible, produces significantly lower DFTs, and seems safe under the co
nditions tested. Biventricular defibrillation may be a useful option for re
ducing DFTs or could be added to an LV pacing lead for heart failure.