Js. Fritz et Sa. Steiner, Effect of a soluble ionic polymer on the separation of anions by capillaryelectrophoresis, J CHROMAT A, 934(1-2), 2001, pp. 87-93
A silica capillary for CE analysis of anions can be conditioned with NaOH,
rinsed with water, coated with a cationic polymer, and equilibrated with th
e background electrolyte in only 2 min for each of the four steps. The coat
ed surface has a positive charge that gives a substantial anodic electroosm
otic flow (EOF) over the range of pH 2.5-12.0. The migration times of sampl
e anions and a neutral marker (used for EOF calculations) are generally rep
roducible to an RSD of 1% or better, both for successive runs on a single c
apillary and from capillary to capillary. It was shown that the type of buf
fer used affects the EOF of a coated capillary. A concentration of 100 mM o
r higher sodium chloride minimizes differences in EOF with different buffer
s and also gives sharper peaks for sample anions. (C) 2001 Elsevier Science
B.V. All rights reserved.