R. Toniolo et al., Amperometric monitoring of hydrogen peroxide in workplace atmospheres by electrodes supported on ion-exchange membranes, J ELEC CHEM, 514(1-2), 2001, pp. 123-128
An amperometric sensor is described which is suitable for the continuous mo
nitoring of traces of hydrogen peroxide present in workplace atmospheres as
gaseous species or as aerosols, without preliminary concentration steps be
ing needed. It consists of a porous Pt sensing electrode, facing the sample
, chemically deposited onto one side of a Nafion(R) 117 membrane which acts
as a solid polymer electrolyte (SPE). The uncoated side of this membrane i
s contacted by an electrolyte solution (0.2 M acetate buffer at pH 4.8, +0.
8 M NaClO4) contained in an internal compartment equipped with a Pt counter
electrode and a Ag \ AgCl \ Cl-(sat.)(-) reference electrode. Hydrogen per
oxide detection was achieved by applying a potential of 1.0 V to the sensin
g electrode, which was chosen on the basis of preliminary voltammetric inve
stigations. The anodic current due to the oxidation of hydrogen peroxide in
creases linearly with its concentration within a wide range (from 0.1 to 40
ppm v/v) and a detection limit of 40 ppb v/v was estimated for a signal-to
-noise ratio of 3. The estimated detection limit is about three orders of m
agnitude lower than I ppm. which is the recommended exposure limit for hydr
ogen peroxide in workplace atmospheres, evaluated as an 8-h time-weighed av
erage concentration. Moreover, the long term stability of this amperometric
sensor appears to be totally satisfactory in that reproducible responses w
ere observed even after long periods of continuous use. The absence of sign
ificant interfering effects is also discussed. (C) 2001 Elsevier Science B.
V. All rights reserved.