An on-line liquid chromatography-immunochemical detection (LC-ICD) system f
or the quantification of cytokines in cell extracts has been developed usin
g a post-column continuous-flow reaction detection system using fluorescenc
e labelled antibodies. Cytokines eluting from the micro-HPLC column react w
ith antibodies to form fluorescent complexes. In a second step the excess o
f free antibody is trapped on a cytokine bound support prior to fluorescenc
e detection. The concentration detection limit of the flow injection-ICD sy
stem was 50 pM (20 mul injection volume) for interleukine 4 (IL-4). An abso
lute detection limit of I fmol was obtained for IL-4. Similar to ICD system
s for small non-protein analytes developed earlier, reaction times were in
the order of I minute. The immobilised cytokine affinity columns can easily
be regenerated and used for months. The present ICD system for interleukin
e 4, 6, 8 and 10 was coupled to ion exchange-, size exclusion- and reversed
phase chromatography. Important parameters (reaction times, reaction condi
tions) were investigated to get a better understanding of post-column ICD s
ystems for macromolecules. (C) 2001 Elsevier Science B.V. All rights reserv
ed.