Ultra-high molecular weight polyethylene (UHMWPE) acetabular cups were test
ed against alumina-ceramic femoral heads using a new type of hip joint simu
lator according to ISO/FDIS 14242-1. Bovine serum as well as newborn calf s
erum were used as test fluids. Total polyethylene wear was determined by we
ight loss of the cups. In addition, wear depth and its distribution were re
corded by means of a coordinate measurement system. Wear particle analysis
and inspection of the worn polyethylene surfaces using light and scanning e
lectron microscopy (SEM) were performed to analyse damage and identify the
acting wear mechanisms.
The total wear rate was determined to be 22.07 +/- 1.75 mg/10(6) cycles for
the bovine serum group and 26.57 +/- 3.55 mg/10(6) cycles for the calf ser
um group. Unexpectedly, the formation of two wear vectors corresponding to
recent clinical findings was detected. Retrieved polyethylene wear debris w
as comparable in size and shape with clinical findings. The test method des
cribed by ISO/FDIS 14242-1 produced reliable and reproducible wear data usi
ng UHMWPE acetabular cups articulating against alumina-ceramic heads. In th
e authors' opinion, the lubricant composition should be described in more d
etail, since the protein and additive content seem to have a high impact on
the wear results. It needs to be emphasized that the findings of this stud
y cannot be regarded as a general validation of hip wear tests according to
ISO/FDIS 14242-1 but are limited to the material combinations investigated
herein. Further testing of other clinically relevant materials and interla
boratory ring tests must follow.