A novel chemiluminescence (CL) sensor for folic acid combined flow-injectio
n (FI) technology was presented in this paper. The analytical reagents invo
lved in the CL reaction, including luminol and hexacyanoferrate(III), were
both immobilized on an anion-exchange column in FI system. The CL signal pr
oduced by the reaction between luminol and hexacyanoferrate(III), which wer
e eluted from the column through sodium phosphate injection, was decreased
in the presence of folic acid, The CL emission was correlated with the foli
c acid concentration in the range from 0.01 to 15 mug ml(-1), and the detec
tion limit was 3.5 ng ml(-1) folic acid (3 sigma). At a flow rate of 2.0 ml
min(-1), including sampling and washing, could be performed in 2 min with
a relative standard deviation of < 2.5%. The flow sensor could be reused mo
re than 300 times and has been applied to the analysis of folic acid in pha
rmaceutical preparations, and the recovery was from 97.4% to 100.4%. (C) 20
01 Elsevier Science B.V. All rights reserved.