A. Coly et Jj. Aaron, Simultaneous determination of binary mixtures of sulfonylurea herbicides in water by first-derivative photochemically induced spectrofluorimetry, J AOAC INT, 84(6), 2001, pp. 1745-1750
First-derivative photochemically induced spectrofluorimetry (PIF-D-1) is ap
plied to the simultaneous determination of binary mixtures of 4 sulfonylure
a herbicides in aqueous micellar samples. Synthetic binary mixtures of sulf
orneturon-methyl with chlorsulfuron, metsulfuron-methyl, and 3-rimsulfuron,
respectively, are well resolved by using the zero-crossing point procedure
. PIF-D-1 allows the determination of binary mixtures of these herbicides w
ith linear dynamic ranges over about 2 orders of magnitude, limits of detec
tion between 0.5 and 52 ng/mL, and relative standard deviations within 0.3-
2.9%. Application to the determination of binary mixtures of these herbicid
es in spiked tap water samples yielded satisfactory recoveries (90-117%).