B. Beneteau-burnat et al., Evaluation of Stratus (R) CS stat fluorimetric analyser for measurement ofcardiac markers troponin I (cTnl), creatine kinase MB (CK-MB), and myoglobin, J CL LAB AN, 15(6), 2001, pp. 314-318
Myoglobin, CK-MB, and Troponin I (cTnl) are cardiac muscle necrosis markers
that are useful for detecting acute myocardial infarction (AMI). The Strat
us (R) CS (Dade Behring, Inc.) is a discrete fluorimetric immunoassay analy
ser designed for the determination of the three cardiac markers from a sing
le sample of whole blood or plasma. Overall analytical performances of the
Stratus (R) CS provided by Dade Behring were evaluated according to the Fre
nch Society of Clinical Biology guidelines. Within-run imprecision (n = 20)
for the three parameters at three levels gave values under 5%, whereas CVs
for between-run imprecision (n = 20) were under 6%. The sensitivities were
0.03 mug/L for cTnl and 0.4 mug/L for CK-MB. Linearities extended from 0-5
0 mug/L for cTnl, 0-140 mug/L for CK-MB, and 1-900 mug/L for myoglobin. The
results, particularly those obtained on whole-blood samples, correlated we
ll with those obtained on Stratus (R) II. We did not find any interference
with haemolysis, icterus, or lipemia. The system was very easy to use, and
fulfills the requirements for the analysis of the three cardiac markers in
patients with acute chest pain in emergency situations. J. Clin. Lab. Anal.
15:314-318, 2001. (C) 2001 Wiley-Liss, Inc.