La. Egnash et R. Ramanathan, Comparison of heterogeneous and homogeneous radioactivity flow detectors for simultaneous profiling and LC-MS/MS characterization of metabolites, J PHARM B, 27(1-2), 2002, pp. 271-284
Methods for simultaneous liquid chromatography-radioactivity monitor (LC-RA
M) metabolite profiling and LC-tandem mass spectrometry (MS/MS) characteriz
ation of metabolites are described. Profiling and characterization of metab
olites from three drug candidates from different therapeutic areas were com
pared using in-line heterogeneous LC-RAM-MS/MS and homogeneous LC-RAM-MS/MS
methods. Although comparison shows that simultaneous metabolite profiling
and characterization can be achieved using either heterogeneous or homogene
ous-LC-RAM-MS/ MS systems, a homogeneous system has the advantage in the fo
llowing aspects, (1) sensitivity; (2) ease of method transfer; (3) less pea
k broadening problems due to the drug or metabolites adhering to the RAM ce
ll; (4) accuracy in quantitation of the metabolites; and (5) the ability to
load larger volumes of unprocessed biological fluids. Furthermore, the stu
dy shows that some of the possible metabolites that do not ionize well with
electrospray ionization (ESI) and eluded detection by heterogeneous-LC-RAM
detection could be very easily detected and characterized using a homogene
ous-LC-RAM-MS/MS system. (C) 2002 Elsevier Science B.V. All rights reserved
.