Jh. Chang et al., An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factor, MED PHYS, 28(11), 2001, pp. 2247-2257
Citations number
22
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
There has been an increasing interest in the application of electronic port
al imaging devices (EPIDs) to dosimetric verification, particularly for int
ensity modulated radiotherapy. Although not water equivalent, the phantom s
catter factor of an EPID, S-pe, is generally assumed to be that of a full p
hantom, S-p, a slab phantom, S-ps or a mini phantom. This assumption may in
troduce errors in absolute dosimetry using EPIDs. A calibration procedure t
hat iteratively updates S-pe and the calibration curve (pixel value to dose
rate) is presented. The EPID (Varian Portal Vision) is irradiated using a
20 x 20 cm(2) field with different beam intensities. The initial guess of d
ose rates in the EPID is calculated from ionization chamber measurements in
air, multiplied by S-p or S-ps. The calibration curve is obtained by fitti
ng EPID readings from pixels near the beam central axis and dose rates in E
PID to a quadratic equation. The S-pe is obtained from EPID measurements in
10 x 10 cm(2) and 20 x 20 cm(2) field and from the calibration curve. and
is in turn used to adjust the dose rate measurements and hence the calibrat
ion curve. The above procedure is repeated until it converges. The final ca
libration curve is used to convert portal dose to dose in the slab phantom,
using the calibrated S-pe, or assuming S-pe = S-p or S-pe = S-ps. The conv
erted doses are then compared with the dose measured using an ionization ch
amber. We also apply this procedure to off-axis points and study its depend
ence on the energy spectrum. The hypothesis testing results (on the 95% sig
nificance level) indicate that systematic errors are introduced when assumi
ng S-pe = S-p or S-pe = S-ps, and the dose calculated using S-pe is more co
nsistent with ionization chamber measurements, Differences between S-pe and
S-ps are as large as 2% for large field sizes. The measured relative dose
profile at d(max) using the EPID agrees well with the measured profile at d
(max) of the isocentric plane using film in a polystyrene phantom with full
buildup and full backup, for open and wedged fields, and for a broad range
of field sizes of interest. The dependence of the EPID response on the ene
rgy spectrum is removed once the calibration is performed under the same co
nditions as the actual measurements. (C) 2001 American Association of Physi
cists in Medicine.