E. Kanehira et al., Development of a new operating rectal tube with a side window for transanal endoscopic surgery, MIN INVAS T, 10(4-5), 2001, pp. 243-247
The authors describe a new rectal tube for use in transanal endoscopic surg
ery. The tube is a transparent plastic sheath with its forward end closed,
measuring 40 mm in diameter and 150 mm in insert length. It has a window, m
easuring 40 mm in diameter, on its side. The target lesion, captured within
the side window, prolapses into the rectal tube and is excellently visuali
sed without positive pressure gas insufflation. This paper gives details of
the tube and reports on a sessile rectal tumour successfully treated using
the new system.