Sources of error, and their correction, in the analysis of isothermal heatconduction microcalorimetric data: applications of a newly developed test reaction
Ak. Hills et al., Sources of error, and their correction, in the analysis of isothermal heatconduction microcalorimetric data: applications of a newly developed test reaction, THERMOC ACT, 380(1), 2001, pp. 19-26
A recent investigation of a test reaction for isothermal heat conduction mi
crocalorimeters, used for the determination of thermodynamic and kinetic pa
rameters, resulted in the following recommended values for the imidazole-ca
talysed hydrolysis of triacetin at 298 K: the reaction rate constant, k = (
2.80 +/- 0.1) X 10(-6) dm(3) mol(-1) s(-1), and the reaction enthalpy chang
e, Delta H-R = -91.7 +/- 3.0 kJ mol(-1). Knowledge of these values allows t
he performance of a microcalorimeter to be evaluated. Thus, it may be possi
ble to correct data for those cases in which there is mis-assignment of the
true instrument baseline zero, of the upper calibration setting and of a d
rift in the instrument baseline. This paper describes the expected data out
put and its analysis in these cases. For these instances, it is possible to
identify the error source and to manipulate subsequent data to produce "co
rrected" results. The procedures are illustrated through the use of data si
mulated as output. (C) 2001 Elsevier Science B.V. All rights reserved.