Direct high-precision measurements of the Sr-87/Sr-86 isotope ratio in natural water, carbonates and related materials by multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS)

Citation
S. Ehrlich et al., Direct high-precision measurements of the Sr-87/Sr-86 isotope ratio in natural water, carbonates and related materials by multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS), J ANAL ATOM, 16(12), 2001, pp. 1389-1392
Citations number
18
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
ISSN journal
02679477 → ACNP
Volume
16
Issue
12
Year of publication
2001
Pages
1389 - 1392
Database
ISI
SICI code
0267-9477(2001)16:12<1389:DHMOTS>2.0.ZU;2-A
Abstract
The present work presents a precise and accurate method for the determinati on of strontium isotope ratios (Sr-87/Sr-86) by multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS). Mass discrimination and instr ument drift are corrected by the use of the natural constant Sr-86/Sr-88 ra tio as an internal standard. The internal standard also corrects for matrix effects, allowing the direct analysis of natural water, carbonates and rel ated materials without chemical separation. The precision (2 sigma) of the Sr-87/Sr-86 ratio is 0.002% (20 ppm), even in samples with total dissolved solid (TDS) levels of up to 500 mg L-1.