Determination of ultra-trace amounts of selenium by continuous flow hydride generation AFS and AAS with collection on gold wire

Authors
Citation
Xm. Guo et Xw. Guo, Determination of ultra-trace amounts of selenium by continuous flow hydride generation AFS and AAS with collection on gold wire, J ANAL ATOM, 16(12), 2001, pp. 1414-1418
Citations number
23
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
ISSN journal
02679477 → ACNP
Volume
16
Issue
12
Year of publication
2001
Pages
1414 - 1418
Database
ISI
SICI code
0267-9477(2001)16:12<1414:DOUAOS>2.0.ZU;2-G
Abstract
This paper describes a method, which is similar to the mercury amalgamation pre-concentration technique, for the determination of ultra-trace amounts of selenium by combining hydride generation and gas-phase enrichment techni ques. The method is based upon the fact that SeH2 can be decomposed to Se-0 and collected by gold wire heated to similar to 200 degreesC. A device for trapping the selenium hydride was designed. On heating the trap at 600 deg reesC or above, the selenium was released rapidly into an atomic fluorescen ce (or AAS) detector as the result of re-forming SeH2 by the reaction betwe en the selenium and the hydrogen carrier gas on the surface of the gold wir e. A detection limit of 5 pg ml(-1) Se for a 5 min collection, and a relati ve standard deviation of 3.0% for 1 ng ml(-1) Se (n = 14) were obtained. Th e method was applied to the determination of selenium in mineral water samp les and materials of a complex nickel-based alloy.