A frequency comparison and an absolute frequency measurement of iodine stab
ilized frequency-doubled Nd:YAG lasers at 532 nm has been performed at the
Max-Planck-Institute for Quantum Optics. Two independent I-2-stabilized las
er systems, one assembled at the Institute of Laser Physics, Novosibirsk, R
ussia, the other at the Physikalisch-Technische Bundesanstalt, Braunschweig
, Germany were investigated. Using a phase-coherent frequency chain, the ab
solute frequency of the I-2-stabilized lasers has been compared to a CH4-st
abilized He-Ne laser at 3.39 mum which has been calibrated against an atomi
c cesium fountain clock. A new value for the R(56)32-0:a(10) component, rec
ommended by the Comite International des Poids et Mesures for the realizati
on of the meter [Metrologia 30 (1993/1994) 523; Metrologia 36 (1999) 211],
has been obtained with reduced uncertainty. Improved absolute frequency val
ues of the R(56)32-0 and P(54)32-0 iodine absorption lines together with th
e hyperfine line separations are presented. (C) 2001 Elsevier Science B.V.
All rights reserved.