Po. Collinson et al., Recalibration of the point-of-care test for CARDIAC T Quantitative with Elecsys Troponin T 3rd generation, CLIN CHIM A, 307(1-2), 2001, pp. 197-203
The rapid troponin T assay CARDIAC T Quantitative was recalibrated using El
ecsys Troponin T 3rd Generation as a new reference method. This paper prese
nts the method comparisons at six centres using the new reference method. M
ethod comparison between CARDIAC T Quantitative Versus Elecsys Troponin T 3
rd Generation were performed using 319 samples from patients with acute cor
onary syndromes. The quality of the CARDIAC T Quantitative was controlled b
y a daily single determination of CARDIAC Control Troponin T, and for the E
lecsys Troponin T 3rd Generation, the Elecsys controls were included in eac
h run. The results for the control materials for the CARDIAC T Quantitative
were between 93% and 107% of the target Values. The CV ranged from 7% to 1
6%. From the regression analysis, according to Bablok and Passing (y = 1.07
x) and the Bland and Altman plot, the bias between CARDIAC T Quantitative a
nd Elecsys Troponin T 3rd Generation is from +6% to +7%, The correlation co
efficient is 0.93, and a 3 x 3 comparison of the clinical efficiency yielde
d 92% clinical concordance between CARDIAC T Quantitative and Elecsys Tropo
nin T 3rd Generation. In conclusion, CARDIAC T Quantitative was in good agr
eement with the reference and calibration method Elecsys Troponin T 3rd Gen
eration. (C) 2001 Published by Elsevier Science B.V.