Cl. Hector et al., The dosimetric consequences of inter-fractional patient movement on three classes of intensity-modulated delivery techniques in breast radiotherapy, RADIOTH ONC, 59(3), 2001, pp. 281-291
Citations number
36
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Onconogenesis & Cancer Research
Background and purpose: A comparison between three classes of intensity-mod
ulated delivery techniques was undertaken to examine the dosimetric consequ
ences of using a multileaf collimator (MLC) reshaped on each imaged fractio
n as opposed to compensators designed on the first day of treatment potenti
ally giving a treatment technique whose accuracy is thus degraded by moveme
nt.
Materials and Methods: The effects of inter-fractional patient movement for
a cohort of six breast patients were studied. Five treatment techniques we
re evaluated, two using a compensator, two using multiple static fields (MS
F) and one using a dynamic multileaf collimator (DMLC). The compensated tec
hniques consisted of (i) the use of compensators designed on day 1 only and
used each fraction thereafter and (ii) the use of a compensator redesigned
for each imaged fraction. The two MSF techniques were (i) a four-field-com
ponent design and (ii) a method where the fluence interval between the MLC
field components was set so they were equivalent to the compensator ('quant
ized' MSF-MLC). The final technique investigated was the DMLC. Plans were p
roduced for each of the five methods and a paired t-test was used to assess
the reduction in the breast volume outside the dose range 95-105% between
sets of pairs of techniques. An on-line correction strategy was simulated t
o determine the number of treatments that required intervention. The action
levels were calculated using the difference between the volume outside the
dose range 95-105% calculated for treatments where the DMLC was designed o
n day 1 only and for each imaged Fraction. Differences of greater than 2%,
greater than 5% and greater than 10% were investigated.
Results: Thirty-five plans were evaluated for each technique. Results showe
d that a statistically significant mean reduction in the volume of the brea
st outside the dose range 95-105% could be achieved if the compensators wer
e designed on each imaged fraction rather than on day 1 only (P = 0.0045).
When the comparison was made between the 'quantized' MSF-MLC and the techni
que where the compensators were designed on day 1 only, a statistically sig
nificant mean reduction in the volume of the breast tissue outside the dose
range 95-105% was not achieved (P = 0.21). Comparison of the DMLC techniqu
e to the technique where the compensators were designed on day 1 only resul
ts in a statistically significant mean reduction in the volume outside the
dose range 95-105% (P = 0.024). This corresponds to a mean reduction in the
volume outside 95-105% dose of 1.94%. The 2% action level showed the great
est reduction in the volume outside 95-105% dose and intervention was only
required in approximately one-third of the treatments investigated.
Conclusions: Redesigning MSFs for each imaged fraction did not provide a st
atistically significant mean reduction in the volume outside the dose range
95-105%. However, using the DMLC technique creates a statistically signifi
cant mean reduction in the volume outside the dose range 95-105%. (C) 2001
Elsevier Science Ireland Ltd. All rights reserved.